4015A¡]500V / 20A¡^AC/DC Digitizing Power
Meter
¡]Includes THD measurement¡^1~4CH
¡]Includes 200Ap electronic
switch + 200A Inrush current measurement + THD
measurement¡^
Description
- 4015A Digital Power Meter is designed to measure the
power line parameters of AC / DC and DC / DC power supply. It is suitable
for AC / DC Adapter, Charger, Quick Charger, LED Driver and DC / DC
Converter. It is also suitable for uninterruptible power supply UPS,
Regulators AVR, Inverter, batteries, AC / DC
power supply / components ... and other applications.
- 4015A digital power meter has three input source
selection switch internally, according to the
test needs to program that one of as input of the power meter.
- Each channel of 4015A digital power meter has a
200Apeak power MOSFET switch parallel relay switch as the output switch, it
can adjustable output on / off and AC any angle,
used to test the Inrush current and Hold up time.
- 4015A digital power meter contains 1 ~ 4 channels
digital power meters, each power meter has a group of high-speed 16-Bits A
/ D Converter, sampling the voltage and current signals respectively.
After the conversion, the data is calculated by a high-speed DSP
processor. After calculation is completed, the data is transmitted to the
host processor and read the relevant information through the RS-232 interface.
- The 4015A has a built-in 4-channel stand alone AC Source / AVR
selector switch and output a 200Apeak, 0 ~ 360º ON /
OFF high-speed switch that turns on the power to be tested one by one and
accurately measures Inrush currents up to 200A peak.
Features
- Suitable for ATE: 500V 20A, 10000W AC/DC Digital
Power Meter¡C
- Voltage Range¡G15 / 30 / 50 / 150 / 300 / 500Vrms
- Current Range¡G0.02 / 0.05 / 0.2 / 0.5 / 2 / 5 /
10 / 20Arms¡B200Apeak¡]for
Inrush¡^
- 0.1% high accuracy
- Measure AC / DC voltage, AC / DC current, AC / DC
power, AC power factor, AC frequency, Inrush current and V/A Harmonic
parameters.
- Measure voltage / current harmonics up to 50th order,
and total harmonic distortion .
- Frequency range DC¡B40~70Hz¡C
- Measuring module: 1 ~ 4CH, standard is 4 channels,
option 1, 2, 3 Channel is available¡C
- The built-in internal low-pass filter (50KHz) can
eliminate unnecessary high-frequency interference
- With AC, DC, AVR 3 sets of
power input, through the control of choice of input power to achieve AC /
DC or DC / DC test flexibility.
- With 0 ~ 360º 200Apeak output switch, can
control the output on / off angle, you can accurately measure the Inrush
current.
Specifications